How do I activate High Sensitivity Mode (HSM) for FLIR G300a?

Atlas SDK

High Sensitivity Mode (HSM) can be enabled and disabled by using the method SetHighSensitivityModeEnabled

Telnet

High Sensitivity Mode (HSM) can be activated through the following procedure when connected through Telnet:

1)

To activate HSM:

rset .image.services.channel.active DIFF

rset .image.services.channel.commit true

2)

To go back to IR:

rset .image.services.channel.active IR

rset .image.services.channel.commit true

3)

To change the HSM level:

rset .image.sysimg.basicImgData.extendedPresentParams.gasDetectDynRange xx

and

rset .image.sysimg.basicImgData.extendedPresentParams.diffImageDelta xx

There are eight different levels of HSM which corresponds to the following resource values:

 Level

 0

 4

 5

 6

 gasDetectDynRange

 6

16 

32 

48 

64 

90 

90 

90 

 diffImageDelta

 0

 0

 0

 0

 0

 0

2

Related Articles