How do I activate High Sensitivity Mode (HSM) for FLIR G300a?
Atlas SDK
High Sensitivity Mode (HSM) can be enabled and disabled by using the method SetHighSensitivityModeEnabled
Telnet
High Sensitivity Mode (HSM) can be activated through the following procedure when connected through Telnet:
1)
To activate HSM:
rset .image.services.channel.active DIFF
rset .image.services.channel.commit true
2)
To go back to IR:
rset .image.services.channel.active IR
rset .image.services.channel.commit true
3)
To change the HSM level:
rset .image.sysimg.basicImgData.extendedPresentParams.gasDetectDynRange xx
and
rset .image.sysimg.basicImgData.extendedPresentParams.diffImageDelta xx
There are eight different levels of HSM which corresponds to the following resource values:
Level |
0 |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
gasDetectDynRange |
6 |
16 |
32 |
48 |
64 |
90 |
90 |
90 |
diffImageDelta |
0 |
0 |
0 |
0 |
0 |
0 |
1 |
2 |